Thanks to the combination of our proven IQ Sensor Net system with the intelligent IDS sensors, the measuring panel delivers accurate and reliable results.

Pre-assembled drinking water panel for monitoring turbidity, pH, conductivity, redox, oxygen and free chlorine.

Each drinking water panel is equipped with piping, pressure reducer, valve, tapping valve, flow indicator and maintenance console.  The version of the measuring panel tailored to your needs also includes the necessary tubing and flow vessels.

Applications:

  • Drinking water
  • Industry
Sensor Measuring range Accuracy Resolution
FCML 412-M12-2 Cl 0.005 … 2.000 mg/l <1 % up to 16 mg/l 0.001 mg/l
FCML 412 M12-20 0.05 … 20.00 mg/l 0.01 mg/l
SenTix® 945 (-P) pH 0.000 … 14.000 pH pH: ±0.004; (IDS measurement electronics)
U [mV]: ±0.2;
T [°C]: ±0.1
0.001 pH;
0.1 mV
SensoLyt® 900- ORP-P
ORP
-1250.0 … +1250.0 mV U [mV]: ±0.2
T [°C]: ±0.1
0.1 mV
TetraCon® 925 (-P) Cond ϰ: 0.0 µS/cm … 2000 mS/cm
SAL: 0.0 … 70.0
TDS: 0 mg/l … 199.9 g/l
±0.5 % from measured value;
T [°C]: ±0.1
ϰ: Up to 0.1 μS/cm
or until 0.01 mS/cm
SAL: 0.1
TDS: Up to 1 mg/l or until 0.01 g/l
LR 925/01(-P) 0.01 … 199.9 µS/cm ±0.5 % from measured value Up to 0.01 µS/cm
Turb PLUS 2000 Turb Factory setting: 0 … 100 NTU; optional: 0 … 10 und 0 … 1000 NTU Up to ±2 % of the measured value
or ±0.02 NTU
Up to 0.0001 NTU
FDO® 925(-P) O2
at 20 °C (68 °F)
0.00 … 20.00 mg/l
0.0 … 200.0 %
±1.5 % 0.01 mg/l
0.1 %
Panel /Measuring system
Outputs Relais:
Power outputs:
Digital:
Number selectable
(mA) Number selectable
Profinet, Profibus, Modbus RTU, Modbus TCP, Ethernet/IP
Data logger Data memory for up to 525,600 data sets
USB interface Included as standard on all panels

Related Products

Color Sensors For Wastewater And Surface Water

Color sensors for wastewater and surface water

Spectral UV-VIS sensor to measure color in wastewater, incl. integrated ultrasonic cleaning; available wavelengths (selectable): 340 nm, 350 nm, 390 nm, 410 nm, 445 nm, 455 nm, 465 nm; including turbidity compensation

Quicktocuv Ii

LAR TOC Analyzer : QuickTOCUV II

Our new QuickTOCuv II is state of the art and has more great features. LAR’s TOC analyzer QuickTOCuv II is ideal for fast and reliable online monitoring in pure, drinking, surface and wastewater.

Carbovis 700 Iq (ts)

CarboVis 700 IQ (TS)

Optimal for the measurement of TOC and DOC in the inlet/effluent of municipal wastewater treatment plants.

Toxalarm

LAR BOD Analyzer : ToxAlarm

LAR’s toxicity analyzer ToxAlarm™ continually monitors drinking and surface water for pollutants. Potential toxins in water are determined through the reaction of highly sensitive bacteria.

Contact Us
By submitting this form with you information, you are indecating of our informal acceptance of our Privacy Policy