FDO700 IQ sensors for dissolved oxygen
Calibration-free, reliable, DIN compliant – the optical FDO® oxygen sensors for the IQ SENSOR NET to regulate biological cleaning steps
Digital ISE sensor for nitrate, connectable to the IQ SENSOR NET
Nitrogen elimination – transparent, process optimized, economical. Nitrate measurement directly in the medium – optimized for regulation purposes:
• Chloride compensation (optional)
• Calibration free, long-term stability
• Economical & cost efficient
With the NitraLyt®Plus 700 IQ, nitrate can be determined directly in the medium. This makes measured values available in a timely manner and they can easily be used for regulation purposes. Another plus of the IQ SENSOR NET probe is the low investment and maintenance cost.
| Measuring method | Electrochemical |
| Appropriate Electrode | Reference electrode VARiON® Ref, Measuring electrode VARiON®Plus NO3, Compensation electrode VARiON®Plus Cl |
| Measuring range/Resolution | NO3-N: 1 … 1000 mg/l / 1 mg/l; 0.1 … 100.0 mg/l / 0.1 mg/l NO3-: 5 … 4500 mg/l / 5 mg/l; 0.5 … 450.0 mg/l / 0.5 mg/l |
| Compensation Range | Cl-: 0.1 … 1,000 mg/l / 0.1 mg/l |
| Measuring Accuracy | ± 5 % of measured value ± 0.2 mg/l in standard solutions |
| Calibration Procedures | Matrix adjustment against any reference value, 2-point-calibration possible with multiple standard solution |
| Working Life (typically) | Reference electrode: 18 months, measuring and compensation electrode: 18 months (in typical application – municipal sewage plants) |
| Temperature Measurement and Compensation | Integrated NTC thermistor, Range 32 °F … 104 °F (0 °C … +40 °C), Accuracy ±0.5 K, Resolution 0.1 K, t95 < 20 s |
| Pressure Resistance | Maximum 0.2 bar (incl. SACIQ sensor connection cable, with installed electrodes) |
| Ambient Conditions | Operating temperature: 32 °F … 104 °F (0 °C … +40 °C), storing temperature: 32 °F … 104 °F (0 °C … +40 °C) |
| Electrical connections | 2-wire shield cable with quick fastener to sensor |
| Electromagnetic Compatibility | EN 61326, Class B, FCC Class A; Intended for indispensable operation |
| Certifications | CE |
| Mechanical | Sensor body: V4A stainless steel 1.4571 Protective cup: POM Temperature sensor: V4A stainless steel 1.4571 Protection rating: IP 68 (0.2 bar, with installed electrodes) Electrode connector: POM |
| Weight | Approx. 1.48 lb (670 g, without electrode, without sensor connection cable) |

Calibration-free, reliable, DIN compliant – the optical FDO® oxygen sensors for the IQ SENSOR NET to regulate biological cleaning steps

LAR’s QuickTONultra continually monitors the TNb content of water. Optionally, further sum parameters can be detected, too. At 1,200 °C, samples are completely oxidised and only within 3 minutes the TNb result is determined.

In contrast to the standard BOD5 method the BioMonitor measures the total BOD. Measurement range: 1-200.000 mg/l BOD, 0 100 % Toxicity, Respiration in mg O2/l x min

LAR’s toxicity analyzer ToxAlarm™ continually monitors drinking and surface water for pollutants. Potential toxins in water are determined through the reaction of highly sensitive bacteria.