IQ optical spectral probes (UV-VIS)
IQ optical spectral probes (UV-VIS) for carbon parameters and suspended solids
Optimal for the measurement of TOC and DOC in the inlet/effluent of municipal wastewater treatment plants.
Digital ISE sensor for ammonium, connectable to the IQ SENSOR NET
Ammonium measurement directly in the medium without sample preparation and sample transfer. Measurement of centrate and other process waters up to 2,000 mg/l NH4-N:
• Calibration free, long-term stability
• No chemicals
• Short response times
The ISE sensor AmmoLyt®Plus 700 IQ allows ammonium measurement directly in the process. Time-consuming steps such as sampling and preparation are eliminated. For stable measured values, the interference substance potassium is compensated simultaneously and can be displayed if needed. An energy-optimized operation by a suitable regulation of the blower aggregates is made possible by this ISE sensor. This will help effectively reduce operating costs of the system, which is amortized by the system in a very short time. Starting now, you will be able to reliably determine centrate and other process water with high ammonium load up to 2.000 mg/l NH4-N with the AmmoLyt®Plus!
| Measuring method | Electrochemical |
| Appropriate Electrode | Reference electrode VARiON® Ref, Measuring electrode VARiON®Plus NO3, Compensation electrode VARiON®Plus Cl |
| Measuring range/Resolution | NH4-N: 1 … 2,000 mg/l / 1 mg/l; 0.1 … 100 mg/l / 0.1 mg/l NH4+: 1 … 2,580 mg/l / 1 mg/l; 0.1 … 129.0 mg/l / 0.1 mg/l |
| Compensation Range | K+: 0.1 … 1,000 mg/l / 0.1 mg/l |
| Measuring Accuracy | ± 5 % of measured value ± 0.2 mg/l in standard solutions |
| Calibration Procedures | Matrix adjustment against any reference value, 2-point-calibration possible with multiple standard solution |
| Working Life (typically) | Reference electrode: 18 months, measuring and compensation electrode: 18 months (in typical application – municipal sewage plants) |
| Temperature Measurement and Compensation | ntegrated NTC thermistor, Range 32 °F … 104 °F (0 °C … +40 °C), Accuracy ±0.5 K, Resolution 0.1 K, t95 < 20 s |
| Pressure Resistance | Maximum 0.2 bar (incl. SACIQ sensor connection cable, with installed electrodes) |
| Ambient Conditions | Operating temperature: 32 °F … 104 °F (0 °C … +40 °C), storing temperature: 32 °F … 104 °F (0 °C … +40 °C) |
| Electrical connections | 2-wire shield cable with quick fastener to sensor |
| Electromagnetic Compatibility | EN 61326, Class B, FCC Class A; Intended for indispensable operation |
| Certifications | CE |
| Mechanical | Sensor body: V4A stainless steel 1.4571 Protective cup: POM Temperature sensor: V4A stainless steel 1.4571 Protection rating: IP 68 (0.2 bar, with installed electrodes) Electrode connector: POM |
| Weight | Approx. 1.48 lb (670 g, without electrode, without sensor connection cable) |

IQ optical spectral probes (UV-VIS) for carbon parameters and suspended solids
Optimal for the measurement of TOC and DOC in the inlet/effluent of municipal wastewater treatment plants.

Spectral UV-VIS sensor to measure color in wastewater, incl. integrated ultrasonic cleaning; available wavelengths (selectable): 340 nm, 350 nm, 390 nm, 410 nm, 445 nm, 455 nm, 465 nm; including turbidity compensation

TetraCon 700 IQ – digital 4 electrode conductivity measuring cell with flow-free operation, especially with high conductivit Very large measuring range from 10 µS/cm … 500 mS/cm

Standard-compliant turbidity measurement with ultrasonic cleaning, according to US EPA 180.1 and DIN EN ISO 7027.