SensoLyt®700 IQ – digital pH/ORP armature with integrated preamplifier and temperature sensor as well as lightning protection to be connected to IQ SENSOR NET:
• Stable signals by digital signal processing
• Convenient by calibration in the lab and glass breakage detection
• Reliable measured values by integrated temperature sensor
By storing calibration values within the sensor, a “pre-calibrated pH-measurement” can be generated. Due to our quick-lock, the sensor can be disconnected and – after laboratory calibration – re-connected at the site easily. Inconvenient calibrations in the field under adverse conditions can be completely eliminated if there is an IQ connection in the lab.
A special switch also allows automatic monitoring for glass breakage. The required replacement of the SensoLyt® electrode is quick and uncomplicated.
The sea water model is optimized for the use in special media: All wetted parts are made of titanium and plastic and are therefore extremely resistant to corrosion.
Modele |
SensoLyt® 700 IQ |
SensoLyt® 700 IQ SW* |
Order No. |
109170 |
109171 |
Measuring method |
Potentiometric |
Measuring range |
0,00 … 14,00 pH (depending on the electrode)
± 2000mV (depending on the electrode) |
Resolution |
0,01 pH
1 mV |
Accuracy |
Depends on calibration
± 0,2 pH; ± 20 mV |
Temp. measurement |
Integrated NTC, -5 °C … +60 °C |
Ambient conditions |
Operating temperature: 0 °C … +60 °C |
Electrical connections |
2-wire shield cable with quick fastener to sensor |
Weight (without cable) |
Approx. 970 g |
Approx. 1.800 g |
* SW: Sensor as sea water model (with plastic arming (POM)) |