IQ SENSOR NET Controller 281

The new System 281 for the IQ Sensor Net – Universal monitor WTW DIQ/S 281-CR2 is a low-cost alternative to single point monitoring of pH / ORP, dissoved oxygen, turbidity or conductivity.

Universal monitor DIQ/S 281-CR2. The new system 281 – the digital and cost-efficient single parameter measuring point with proven IQ Sensor Net technology and IQ sensors for the standard parameters. Low-cost alternative to analog measuring points
The System 281 combines the known performance of analog products with the advantages of the digital IQ Sensor Net:
• Proven and reliable IQ Sensor Net technology
• 100 % IQ Sensor Net look & feel
• 2 mA for main and secondary parameter
• 2 relays for alarm for main and secondary value
• You can easily upgrade to DIQ/S 282/284, since IQ sensors can be retained
• Free software updates with new features
• Matching digital IQ sensors for pH/ORP, conductivity, O2 and turbidity
• Stable, robust and proven measuring technology
With the system 281, WTW offers a digital and cost-effective alternative to analog measuring points. Our IQ sensors (also as seawater version) are available for the parameters pH/redox, conductivity, oxygen or turbidity.
You will receive a cost-effective measuring point with proven WTW IQ Sensor Net technology. In addition to the power supply via a wide range power supply (100-240VAC) or via 24 V, 2 current outputs and 2 relays are also available.

DIQ/S 281- CR2(/24V) MOD HART
Connectable sensors IQ sensors for pH/ORP, O2, conductivity, turbidity, TSS or sludge level
Current outputs and relays 2 x (0) 4–20 mA
2 x relays
Modbus RTU
2 x relays
HART
2 x relays
Cable length sensor 100 m fixed cable
Max. total cable length 250 m
Power supply Wide range mains converter (100-240VAC)
or 24V
Connectable modules DIQ/CHV (Cleaning Head Valve), DIQ/JB and
MIQ/WL PS SET

Related Products

LAR TOC Analyzer : QuickTOC purity

The QuickTOCpurity from LAR Process Analysers AG is the online TOC measurement system for the determination of total carbon (TC), total organic carbon (TOC), total inorganic carbon (TIC) and dissolved organic carbon (DOC) in pure water.

LAR BOD Analyzer : BioMonitor

In contrast to the standard BOD5 method the BioMonitor measures the total BOD. Measurement range: 1-200.000 mg/l BOD, 0 100 % Toxicity, Respiration in mg O2/l x min

UV sensors for nitrate

UV 700 IQ NOx: Low-cost probe with integrated ultrasonic cleaning for the maintenance-free and reagent-free measurement of nitrate

Contact Us
By submitting this form with you information, you are indecating of our informal acceptance of our Privacy Policy