UV701 UV IQ sensors for SAC, 1 mm
UV705 UV IQ sensors for SAC, 5 mm
SAC and UVT sensor for the IQ SENSOR NET
UV 700 IQ SAC: Low-cost probe (integrated ultrasonic cleaning, turbidity compensation) for the maintenance-free and reagent-free SAC measurement according to DIN 38404 C3:
• No wear parts
• No replacement of seals
• No risk of penetrating water
• No corrosion
With the new sensor UV 700 IQ SAC, you can determine the spectral absorption coefficient at 254 nm directly and without chemicals. Turbidity influences are compensated by a reference measurement at 550 nm.
As a sum parameter, similar to the COD, the SAC allows information about the organic contamination of water. This makes this sensor ideal to capture organic deposits in wastewater and surface waters. In addition to the SAC, you can use this sensor to determine the UV transmission at 254 nm. This is most commonly used in the control of UV disinfection systems.Specifications

  Model UV 701 IQ SAC   UV 705 IQ SAC
  Measuring method UV-Absorptionsmessung 254 nm (Kompensation 550 nm)
  Measuring gap (optical layer thickness) 1 mm 5 mm
  Application(optimized for) Municipal wastewater with a low proportion of industrial wastewater,
wastewater treatment plants, surface water
Measuring range and Resolution  COD   0.0 … 12,500 mg/l 1 mg/l   0.0 … 800 mg/l 1 mg/l
 TOC   0.0 … 20,000 mg/l 1 mg/l   0.0 … 500.0 mg/l 0.1 mg/l
 DOC   0.0 … 12,500 mg/l 1 mg/l   0.0 … 500.0 mg/l 0.1 mg/l
 BOD   0.0 … 8,000 mg/l 1 mg/l   0.0 … 500.0 mg/l 0.1 mg/l
 SAC254 total   0.0 … 3,000 1/m 1 1/m   0.0 … 600.0 1/m 0.1 1/m
 SAC254 dissolv   0.0 … 3,000 1/m 1 1/m   0.0 … 600.0 1/m 0.1 1/m
 UVT254 total*   0.0 … 100.0 % 0.1 %   0.0 … 100.0 % 0.1 %
 UVT254 dissolv*   0.0 … 100.0 % 0.1 %   0.0 … 100.0 % 0.1 %
 Flow rate   ≤ 3 m/s
 Pressure Resistance   Maximum 1 bar (incl. sensor connection cable)
 Electrical connections   2-wire shield cable with quick fastener to sensor
 Electromagnetic Compatibility    EN 61326, Class B, FCC Class A
  Intended for indispensable operation
 Certifications   CE
 Mechanical   Housing: Titan Grade 2, PEEK
  Window: Sapphire glass
  Protection class: IP 68
 Weight (without cable)   Approx. 8.82 lb (4 kg)
 Warranty   2 years for defects in quality
* The UVT-254 value is standardized to 10 mm gap width.

Related Products

IQ suspended solids sensor

ViSolid®: Optical sensors for the in-situ use to measure suspended solids via scattered light and direct back-scattering with ultrasonic cleaning system

LAR TOC Analyzer : QuickTOC uv

QuickTOCuv™ is an online measurement system for the determination of total organic carbon (TOC) and the dissolved organic carbon (DOC) in pure, drinking, surface and wastewater.

LAR TOC Analyzer : QuickTOCeco

LAR’s QuickTOCeco™ is an online analyzer for the Total Organic Carbon (TOC) in water and process water, as condensate return and cooling water. At 1,200°C

Contact Us
By submitting this form with you information, you are indecating of our informal acceptance of our Privacy Policy