Product Descriptions
This wall- mounted instrument is the most versatile remote air quality monitor available for fixed installations. The advanced power management and the industry’s first sleep mode, allows for long intervals between samples over extended and unattended operations. The AQM Series can easily integrate into a building automation and facility monitoring system via Ethernet, USB or (optional) Wireless 802.11 b/g, RS485 or RS232.
Models |
5301-AQM1 and 5301-AQM2 |
Size Range |
0.3 to 25 μm |
Size Channels |
Factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 μm variable binning |
Flow Rate / Accuracy |
0.1 CFM (2.83 LPM) / +/- 3% |
Aerosol Concentration Range |
0.01 to 20,000 μg/m³ |
Light Source |
Long life laser diode |
Resolution |
.00003 μg/m3 |
Zero Stability |
No appreciable drift |
Count Modes |
Real-Time Meter and graph, cumulative/differential count/m³ and count/ft³, and mass concentration (PM) |
Log Interval |
Minimum 1 second with no hold time, maximum 99 hours |
Calibration |
NIST traceable |
Display |
4.3″ (10.9 cm) WQVGA (480×272) color touch screen |
Printer |
(Optional) External thermal printer available |
Internal Vacuum Pump |
Internal pump with automatic flow control |
Filtered Exhaust |
Internal HEPA filter |
Number of Channels |
6 Calibration for custom size channels available |
Custom Size Channels |
Calibration for custom size channels available |
Audible Alarm |
Adjustable built-in alarm |
Communication Modes |
Ethernet, USB. (Optional) Wireless 802.11 b/g, RS485 or RS232. |
Environmental Sensors 5301-AQM |
Includes NDIR CO2 (0-5000 ppm, resolution 1 ppm, accuracy ±1% FS, Response rate 20 seconds), temperature and relative humidity probe 32º to 122ºF (0º to 50ºC) ±1ºF (0.5ºC), 15-90% ±2% RH |
Environmental Sensors 5302-AQM |
Includes all sensors in the 5301-AQM1 and TVOC PID (standard on the 5301-AQM2), 0-50 ppm / minimum detection level 5 ppb, accuracy ±1.5%, Response rate <3 seconds |
Alarm |
Alarms on counts for all particle sizes, low battery, sensor failure, environmental sensors and flow |
Standards |
ISO 21501-4 and JIS B9921 |
Instrument Calibration |
Recommended minimum once per year. Gases, temperature and humidity sensors field calibratable. |
External Surface |
Stainless steel |
Dimensions (L x W x H) |
5.22” x 2.25” x 9.13” (13.3 cm x 5.7 cm x 23.2 cm) includes barb fittings |
Weight |
4 lb (1.8 kg) |
Accessories |
Operating Manual and IMS Software on USB flash drive, isokinetic probe, temperature relative humidity sensor, purge filter, battery, USB cable, and power supply |
Optional Accessories |
Calibration cap for gas sensor, carrying case, external printer, isokinetic probes, barb fittings, spare battery, external battery charger, printed manual, and IMS-RT monitoring system |
Data Storage |
45,000 sample records (rotating buffer) includes particle count data, environmental data, locations, times and annotations, scrollable on screen or printout |
Sample Time |
1 second to 99 hours |
Power |
110 to 240 VAC 50/60 Hz universal in-line power supply |
Operating Conditions |
41º to 104ºF (5º to 40ºC) / 20% to 95% non-condensing |
Storage Conditions |
32º to 122ºF (0º to 50ºC) / Up to 98% non-condensing |
Warranty |
2 year limited warranty. Extended warranties available. |