Digital electro-chemical IQ sensors for dissolved oxygen, connectable to the IQ SENSOR NET

IQ pH/ORP armatures
Digital pH/ORP
Digital electro-chemical IQ sensors for dissolved oxygen, connectable to the IQ SENSOR NET
TriOxmatic® IQ: Reliable and proven digital electro-chemical oxygen sensors with 3 electrode system (ECDO) for precise and accurate measurements:
• Low investment costs
• No running in, no long-term drift – stable from the beginning to the end
• Self-diagnosis system SensReg/SensLeck by means of 3 electrode system
The electro-chemical sensors are known for their high accuracy – from the beginning. Their teflon membrane makes them extremely insensitive to biological deposits. Therefore, in most cases, you will not need an additional cleaning unit.
Also, the membrane – just like the electrolyte consumption – is monitored continuously. Our self-diagnosis system SensLeck and SensReg show occurring errors automatically.
The sea water model is optimized for the use in special media: All wetted parts consist of titanium and plastic and are therefore extremely corrosion-resistant.
Furthermore, if you want, you will find a sensor with a faster response time or for the measurement of trace oxygen.
| Model | TriOxmatic® 700 IQ | TriOxmatic® 700 IQ SW* | TriOxmatic® 701 IQ | TriOxmatic® 702 IQ |
| Order No. | 201640 | 201641 | 201644 | 201646 |
| Measuring method | Electrochemical | |||
| Measuring range (25 °C) | ||||
| O2 concentration | 0.0 … 60.0 mg/l | 0.00 … 20.00 mg/l 0.0 … 60.0 mg/l |
0 … 2000 µg/l 0.00 … 10.00 mg/l |
|
| O2 saturation | 0 … 600% | 0.0 … 200.0% 0 … 600% |
0 … 110% | |
| Resolution | ||||
| O2 concentration | 0.1 mg/l | 0.01 mg/l 0.1 mg/l |
0.001 mg/l 0.01 mg/l |
|
| O2 saturation | 1% | 0.1% 1% |
0.1% | |
| Response time at 25 °C | t90: 180 s | t90: 30 s t99: 90 s |
t90: 30 s t99: 110 s |
|
| Minimum flow rate | 0.05 m/s | 0.23 m/s | 0.3 m/s | |
| SensCheck | SensLeck SensReg |
SensReg | SensLeck SensReg |
— SensReg |
| Temp. measurement | Integrated NTC, 23 °F … 140 °F (-5 °C … +60 °C) ± 0.5 °C | |||
| Temp. compensation | 32 °F … 140 °F (0 °C … +60 °C) | |||
| Pressure Resistance | 10 bar (incl. sensor connection cable) | |||
| Ambient Conditions | Operating temperature: 32 °F … 140 °F (0 °C … +60 °C); Storage temperature: 23 °F … 149 °F (-5 °C … +65 °C) |
|||
| Electrical connections | 2-wire shield cable with quick fastener to sensor | |||
| Electromagnetic Compatibility | EN 61326, Class B, FCC Class A; Intended for indispensable operation |
|||
| Certifications | CE, cETL, ETL | |||
| Mechanical | Membrane head assembly, locking cap: POM Sensor body: V4A stainless steel 1.4571 Protection rating: IP 68 |
|||
| Weight (without cable) | Approx. 1.46 lb (660 g) | Approx. 2.58 lb (1,170 g) | Approx. 1.46 lb (660 g) | |
| * SW: Sensor as sea water model (with plastic arming (POM)) | ||||

Digital pH/ORP

Smart signal processing – the sensor has a special pattern detection feature and can therefore distinguish between the sludge level and undesired signals.

LAR’s toxicity analyzer NitriTox™ continually monitors water treatment plants for pollutants. Potential toxins in water are determined through the reaction of highly sensitive bacteria.

UV 700 IQ NOx: Low-cost probe with integrated ultrasonic cleaning for the maintenance-free and reagent-free measurement of nitrate