IQ SENSOR NET Controller 281
Universal monitor DIQ/S 281-CR2. The new system 281 – the digital and cost-efficient single parameter measuring point with proven IQ Sensor Net technology and IQ sensors for the standard parameters.
LAR’s QuickTOCeffluent is the ultimate online TOC analyzer for the determination of total organic carbon (TOC) for waste water treatment plants effluents. Products by Brand LAR, Germany.
QuickTOCeffluent
Features & Benefits
• Measurement range: 0.1 – 200 mg/l
• Response time: 2-3 min
• TC, TOC monitoring
• Easy calibration and validation (QuickCalibration)
• Highest combustion temperature (1,200°C) on the market
• Catalyst-free technique
• No filtration necessary at the effluent
• No memory effects
• Low operational and maintenance costs
• According to DIN EN 1484:1997-08, ISO 8245:1999-03 and EPA 415.1
Applications
LAR’s total organic carbon analyzer is suitable for a variety of applications and types of water:
• Municipal WWTP outlet
• Industrial WWTP outlet
• Process water
Measurement and Sample Preparation | |
Type of measurement: thermal combustion at 1,200 | |
Measuring range | 0.1 – 200 mg/l (ppm) |
Response time | 2-3 minutes (application dependent |
Reproducibility | ± 2% |
Accuracy | ± 2% |
Connections | |
Sample water, in | Prene tube 3.2 x 1.6mm |
Sample water, drain | PVC tube 12 x 2mm |
Electrical power | ~115 / 230V, 50 / 60 Hz |
Analog output | 0/4 – 20 mA |
Serial interface | RS 232 for remote control Malfunction alarm, life-zero |
Status output | 4 relais contacts (programmable) |
Remote control | via TCP/ IP protocol (internet) |
Dimensions and Weights | |
Cabinet | steel IP 54 |
Option | stainless steel, IP 65, ATEX zone 1 and zone 2 |
Dimensions | 1,060 x 600 x 520 mm (H x W x D) |
Weight | approx. 115 kg (254 lb) |
Universal monitor DIQ/S 281-CR2. The new system 281 – the digital and cost-efficient single parameter measuring point with proven IQ Sensor Net technology and IQ sensors for the standard parameters.
UV 700 IQ NOx: Low-cost probe with integrated ultrasonic cleaning for the maintenance-free and reagent-free measurement of nitrate
ViSolid®: Optical sensors for the in-situ use to measure suspended solids via scattered light and direct back-scattering with ultrasonic cleaning system
Spectral UV-VIS sensor to measure color in wastewater, incl. integrated ultrasonic cleaning; available wavelengths (selectable): 340 nm, 350 nm, 390 nm, 410 nm, 445 nm, 455 nm, 465 nm; including turbidity compensation